SETPOINT 2 CONFIGURATION (S2.CF)

Setpoint 2 Configuration ("S2.CF") is not active unless your meter has dual relay output capabilities. The LED's will display whether the ("S2.CF") is active or not. You may use Setpoint 2 Configuration ("S2.CF") for the following:
    * To set setpoint's active band above or below your chosen value
    * To select whether the setpoint operation is latched or unlatched
1. Press the MENU button until the meter shows "S2.CF".

2. Press the /DEV button. The meter shows one of the following:
    * "S.1=A" (Active above the setpoint) (default)
    * "S.1=B" (Active below the setpoint)
3. Press the /MAX button to toggle between available choices.

4. Press the /DEV button again The meter shows one of the following:
    * "S.2=L" setpoint 1 to be latched
    * "S.2=U" setpoint 1 to be unlatched
5. Press the /MAX button to toggle between available choices.

6. Press the MENU button to store your selection(s). The meter momentarily shows "STRD", followed by "S1.DB" (Setpoint 1 Deadband).

SETPOINT 1 DEADBAND (S1.DB)

Setpoint 1 Deadband is not active unless your meter has dual relay output capabilities. The LED's will display whether the ("S1.DB") is active or not. To set the deadband (hysteresis) of setpoint 1, follow these steps:

1. Press the MENU button until the meter shows "S1.DB".

2. Press the /DEV button. The meter shows the last previously stored 4-digit number (0000 through 9999) with flashing 4th digit.

3. Press the /MAX button to change the value of the flashing digit. If you continue to press the /MAX button, the flashing digit's value continues to change.

4. Press the /DEV button to scroll horizontally to the next digit.

5. Press the MENU button to store value. The meter momentarily shows "STRD", followed by "S2.DB" (Setpoint 2 Deadband).


SETPOINT 2 DEADBAND ("S2.DB")

Setpoint 2 Deadband ("S2.DB") is not active unless your meter has dual relay output capabilities. The LED's will display whether the (S2.DB) is active or not. To set the deadband (hysteresis) of setpoint 2, follow these steps:

1. Press the MENU button until the meter shows "S2.DB".

2. Press the /DEV button. The meter shows the last previously stored 4-digit number (0000 through 9999) with flashing 4th digit.

3. Press the /MAX button to change the value of the flashing digit. If you continue to press the /MAX button, the flashing digit's value continues to change.

4.
Press the /DEV button to scroll to the next digit.

5. Press the MENU button to store your selection.
"STRD" momentarily displays, followed by cold junction offset if you have a standard meter (refer to Section 19) or "OT.CF" (Output Configuration) if you have analog output capabilities (refer to Section 15).